DISPlay Subsystem

DISPlay[:WINDow[1]|2]:SELect[1]|2

This command is used to select a specific measurement within a specific window.
If the second numeric value is not sent, the upper measurement of the relevant
window is selected. This command is used to specify which measurement is used
for the analog, trace, or single numeric display.
Syntax
Example
DISP:WIND2:SEL1
Reset Condition
On reset, the upper window upper measurement is selected.
Query
DISPlay[:WINDow[1]|2]:SELect[1]|2?
The query enters a 1 or 0 into the output buffer indicating whether the window
specified is currently selected.
– 1 is returned if the specified window is selected
– 0 is returned if the specified window is not selected
254
DISP
:WIND
1
:SEL
2
This command selects the upper
measurement in the lower window.
N1911A/1912A P-Series Power Meters Programming Guide