SENSe Subsystem
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed] <numeric_value>
This command is used to enter a frequency. If the frequency does not correspond
directly to a frequency in the sensor calibration table, the power meter calculates
the calibration factor using linear interpolation. For 8480 Series power sensor the
power meter uses linear interpolation to calculate the calibration factor for the
frequency entered if [SENSe[1]]|SENSe2:CORRection:CSET:STATe is ON. For
P-Series and E-Series power sensor, the appropriate corrections are applied for
the frequency selected, dependant on the calibration data stored in the sensor's
EEPROM.