Figure 1-8 Typical Status Register Bit Changes - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

Enable Register
The enable register specifies the bits in the event register that can generate a
summary bit. The instrument logically ANDs corresponding bits in the event and
enable registers and ORs all the resulting bits to obtain a summary bit. Enable
registers are read-write. Querying an enable register does not affect it.
An Example Sequence
Figure 1-8
for various settings. The changing state of the condition in question is shown at
the bottom of the figure. A small binary table shows the state of the chosen bit in
each status register at the selected times T1 to T5.
A
B
C
D
Figure 1-8
N1911A/1912A P-Series Power Meters Programming Guide
illustrates the response of a single bit position in a typical status group
0
0
0
0 0
0
0 0
0 1
1
1 0
1
0 0 0
1
1
0
0 0 0
1
Condition
0
T1
*
Typical Status Register Bit Changes
Power Meter Remote Operation
0
0
0
1
1
0
1
1
0
1
1
1 0
1
1
1
0
1
T2
*
*
marks when event register is read
0
0
0 0
0
0
0
0 1
1
0
0
0
0
0
0
0
1
0
T3
T4
1
0 0
0
0
0 0
0
0
0
0
0
0
*
T5
75