[Sense[1]]|Sense2:Frequency[:Cw|:Fixed]:Step <Numeric_Value - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXEd]:STEP
<numeric_value>
This command sets the number of steps in average or peak frequency sweep. It
must be used in conjunction with external trigger.
The frequency sweep range will be equally divided by the frequency step. If trace
display is turned on, the measurement window will be restored to single numeric
or analog depends on the number of measurement channel.
Determine the Right Step to be Set
NOTE
Number of frequency step can be calculated using equation below:
Step = f
where,
Step = Number of frequency step
f
start
f
stop
Interval = Frequency step size
Example
When f
should be set to
Step = f
N1911A/1912A P-Series Power Meters Programming Guide
– f
+ Interval
stop
start
Interval
= Frequency sweep's start point
= Frequency sweep's stop point
= 1 GHz and f
start
– f
+ Interval
stop
start
Interval
= 5 GHz – 1 GHz + 0.5 GHz
0.5 GHz
= 9
= 5 GHz with given interval of 0.5 GHz, the Step
stop
SENSe Subsystem
10
421