Trace[1]|2:Measurement:reference? <Numeric_Value - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

13
TRACe Subsystem
TRACe[1]|2:MEASurement:REFerence? <numeric_value>
This command is used to find the reference power level. This provides the
reference power level to calculate the pulse parameters.
Commonly used reference levels are 0 %, 10 %, 50 %, 90 %, and 100 %. You can
set the reference level to measure overshoot at 125 % and undershoot at –25 %.
Algorithm
P
x%
where:
– 0 % <= x <= 100 %
– P
– P
– P
Syntax
Reset Condition
On reset, this parameter is not affected.
Example
632
= P
+ x/100 (P
0%
100%
= level of low state
0%
= level of high state
100%
, P
and P
0%
100%
x%
Watts.
TRAC
1
2
TRAC2:MEAS:REF? 100
N1911A/1912A P-Series Power Meters Programming Guide
- P
)
0%
are all in the same unit of measurement, for example,
:REF
:MEAS
This command returns the high state
power for trace 2.
numeric_value
?