Item
resolution
source list
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command sub-systems. The parameters must
be entered in the specified order. If parameters are omitted, they default from the right. The parameter DEFault is used as a place holder.
Specifying DEF leaves the parameter value unchanged.
[b] When the measurement result is linear this parameter represents the number of significant digits. When the measurement result is
logarithmic 1 to 4 represents of 1, 0.1, 0.01 and 0.001 respectively.
Example
READ2:POW:AC?
Error Messages
– INITiate:CONTinuous must be set to OFF, otherwise error –213, "INIT ignored"
– If TRIGger:SOURce is set to BUS or HOLD, error –214, "Trigger deadlock"
– If the expected value and resolution parameters are not the same as the
N1911A/1912A P-Series Power Meters Programming Guide
Description/Defaul t
A numeric value for the resolution. If it is unspecified the
current resolution setting is used. If a value is entered it
should correspond to the current resolution setting
otherwise an error occurs.
The channel which the command is implemented on.
If unspecified the current window setup is used. However,
on the N1912A, if the window shows a ratio or difference
measurement, the upper window defaults to Channel A and
the lower window to Channel B.
occurs.
occurs.
current expected value and resolution settings on the specified window, error
–221, "Settings conflict" occurs.
MEASurement Commands
Range of Values
1 to 4
1.0, 0.1, 0.01, 0.001
DEF
(@1)
(@2) (N1912A only)
This command queries the lower window/
upper measurement.
[b]
[a]