[Sense[1]]|Sense2:Buffer:count <Numeric_Value - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

10
SENSe Subsystem
[SENSe[1]]|SENSe2:BUFFer:COUNt <numeric_value>
This command sets the buffer size for average or peak measurement. It must be
used in conjunction with external trigger.
It can only be set when frequency sweep is disabled (FREQ:STEP 0). Otherwise,
this parameter will be automatically overwritten by frequency sweep step. If trace
display is turned on, the measurement window will be restored to single numeric
or analog depends on the number of measurement channel.
This command is only applicable when used with 8480, N8480, E4410, E9300,
NOTE
E9320 or N1920 power sensors (average or peak).
Syntax
SENS
SENS2
Parameters
Item
numeric_value
388
:
FREQ
1
Description
A numeric value for buffer size.
N1911A/1912A P-Series Power Meters Programming Guide
:COUN
Space
numeric_value
?
Range of Values
1 to 2048