CALibration[1]|2:RCFactor <numeric_value>
This command is used with 8480 Series power sensors or N8480 Series power
sensors with Option CFT to set the reference calibration factor of the specified
channel. Reference calibration factors can also be set using sensor calibration
tables. The power meter uses the most recently set reference calibration factor.
Syntax
Parameters
Item
numeric_value
N1911A/1912A P-Series Power Meters Programming Guide
CAL
1
2
Description/Defaul t
A numeric value:
– DEF: the default is 100 %
– MIN: 1 %
– MAX:150 %
Space
:RCF
?
CALibration Subsystem
numeric_value
DEF
MIN
MAX
Space
MIN
MAX
Range of Values
1.0 to 150.0 PCT
DEF
MIN
MAX
4
223