Command Reference
E4981A SCPI Commands

:SYSTem:COMMunicate:LAN[:SELF]:CONFigure

Syntax
:SYSTem:COMMunicate:LAN[:SELF]:CONFigure {AUTO|MANual}
:SYSTem:COMMunicate:LAN[:SELF]:CONFigure?
Description
Sets the IP configuration setup method i.e. Auto/Manual.
Parameters
AUTO (Initial [Factory set]
value)
MANual
The value of this command is not changed with
and is only changed with Factory Default reset (available only through
Front Panel).
{AUTO|MAN}<newline><^END>
Query response
Equivalent key
[System] - SYSTEM CONFIG - IP CONFIG - AUTO|MANUAL
sequence

:SYSTem:COMMunicate:LAN[:SELF]:CONTrol?

Syntax
:SYSTem:COMMunicate:LAN[:SELF]:CONTrol?
Description
Returns the SOCKET control port number. If the parser is a SOCKET, it will
return a number from 5000 to 5100. Otherwise, 0. (Query Only)
Query response
{numeric}<newline><^END>
Equivalent key
No equivalent key is available on the front panel.

:SYSTem:COMMunicate:LAN[:SELF]:CURRent:ADDRess?

Syntax
:SYSTem:COMMunicate:LAN[:SELF]:CURRent:ADDRess?
Description
Returns the current IP address. (Query Only)
Query response
{"string"}<newline><^END>
Equivalent key
No equivalent key is available on the front panel.
276
Description
Sets the IP configuration setup to auto mode.
Sets the IP configuration setup to manual mode.