3
CALCulate Subsystem
CALCulate[1]|2|3|4:LIMit:LOWer[:DATA] <numeric_value>
This command enters a value for the lower test limit for the specified window/
measurement used in the CALCulate[1]|2|3|4:LIMit:FAIL? test. The units used
are dependent on the current setting of
UNIT:POWer and CALCulate:RELative:STATe as shown in
measured value is less than the value specified in
CALCulate[1]|2|3|4:LIMit:LOWer[:DATA], CALCulate[1]|2|3|4:LIMit:FAIL?
reports a fail. When the measured value is greater than or equal to the limit, a fail
is not reported.
Table 3-1
Measurement Units
Measurement Mode
Single Channel
Ratio
Difference
Syntax
CALC
194
Measurement Type
Avg, Pk
Pk–Avg
Avg, Pk, Pk–Avg
Avg, Pk
Pk–Avg
:LIM
1
2
3
4
N1911A/1912A P-Series Power Meters Programming Guide
CALC:REL:STAT OFF
Linear
Watt
%
%
Watt
%
:LOW
:DATA
Table
CALC:REL:STAT ON
Log
Linear
dBm
%
dB
%
dB
%
dBm
%
dB
%
Space
numeric_value
DEF
MIN
MAX
Space
?
3-1. When the
Log
dB
dB
dB
dB
dB
MIN
MAX