10
SENSe Subsystem
Reset Condition
On reset, the filter length is set to 4.
Query
AVERage2:COUNt?
The query returns the current setting of the video filter length. The format of the
response is <NR1>.
Query Example
AVER2:COUN?
Error Messages
– If the command is used when a sensor other than a P-Series or E9320 power
– If the command is used when an E9320 sensor is connected and set to
382
sensor is connected, error –241, "Hardware missing" occurs
AVERage mode rather than NORMal mode, the error –221, "Settings Conflict"
occurs.
N1911A/1912A P-Series Power Meters Programming Guide
This command queries the video filter length
for Channel A.