10
SENSe Subsystem
[SENSe[1]]|SENSe2:BUFFer:MTYPe <string>
This command sets the measurement type to be returned from the buffer. It can
only be used in conjunction with external trigger.
This command is only applicable when connecting E9320 or N1920 sensor in
NOTE
peak mode.
Syntax
Parameters
Item
string
390
:
SENS
1
2
3
4
SENS2
Description
The input measurement type to be fed to the specific input on the
SENSe block:
– PEAK: peak power
– PTAV: peak to average
– AVER: average
– MIN: minimum power
N1911A/1912A P-Series Power Meters Programming Guide
:MTYPe
FREQ
string
Space
?
Range of Values
"PEAK"
"PTAV"
"AVER"
"MIN"