Service:sensor[1]|2:Pcalfactor <Cal_Factor_Data - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

SERVice:SENSor[1]|2:PCALfactor <cal_factor_data>
This command writes calibration factor data to, or reads calibration factor data
from, the currently connected sensor. The whole calibration factor block must be
written at once as a checksum is generated. The new block must not be larger
than the existing block.
This command applies to E9320 Series sensors for peak path data only. For E4410
Series, E9300 Series and E9320 Series sensors, average path data, refer to
"SERVice:SENSor[1]|2:CALFactor <cal_factor_data>"
Syntax
SERV
Parameters
Item
cal_factor_data
Query
SERVice:SENSor[1]|2:PCALfactor?
The query returns the current peak path calibration factor block.
N1911A/1912A P-Series Power Meters Programming Guide
:SENS
1
2
Description/Defaul t
A binary data block. Refer to
further information.
Space
:PCAL
Appendix A, "Calibration Factor Block Layout,"
SERVice Subsystem
on page 705.
cal_factor_data
?
on page 759 for
15
713