Trace[1]|2:Measurement:pulse[1] - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

TRACe[1]|2:MEASurement:PULSe[1]|...|10:DCYCle?
This command returns the duty cycle of the selected pulse in percentage.
Algorithm
Duty Cycle = (pulse duration / pulse period) * 100
where,
pulse duration is the time difference between positive and negative transitions of
one pulse and pulse period is the time difference between two consecutive
transition occurrences of the same polarity.
Syntax
Example
Error Messages
– The command is only applicable when N192x or E932x Sensors are present,
– If free run acquisition is selected or sensor average mode selected, Error –221
If you attempt to measure a pulse out of the range of the capture, for example,
NOTE
measure the 5th pulse and there are only 4 pulses displayed, the power meter
returns #0##9.91E37 as the result.
N1911A/1912A P-Series Power Meters Programming Guide
TRAC
1
2
TRAC2:MEAS:PULS3:DCYC?
otherwise Error –241, "Hardware Missing" is generated.
"Settings Conflict" occurs.
:PULS
:MEAS
This command returns the duty cycle of the
3rd pulse found on trace 2.
TRACe Subsystem
1
:DCYC
2|...|10
13
?
621