1.9 Configuring X-ray components via the SIDEXIS Manager

59 38 399 D3352
D3352.076.01.26.02 09.2016
1.9
Configuring X-ray components via the
SIDEXIS Manager
1.9.1

Setting up the X-ray component

Since it is addressable via the network, the X-ray component can in principle
be activated for X-ray image acquisition by any of the PCs connected to the
network.
The SiXABCon program is used to manage networkable X-ray components.
It can be executed on any PC in the network, on which SIDEXIS XG has been
installed.
IMPORTANT! In order to avoid IP address conflicts, you should never oper-
ate several networkable X-ray components in the network using the same IP
address. Each X-ray component needs to be assigned a unique IP address.
Factory setting of the TCP/IP address of the unit:
192.168.15.240 (subnet mask: 255.255.255.0)
Checking existing IP addresses
To find out whether an IP address already exists in the network, enter the
"PING" function in the input prompt (DOS window).
1.
Switch on all network devices (computers, printers, X-ray components)
which have been operated to date.
2.
Invoke the input prompt (DOS window) from a network computer.
3.
At the input prompt, enter "ping" followed by the address to be checked
and then press the Enter key.
Example: "ping 192.168.15.13"
– If a network device responds, then this address has already been as-
signed.
1 – 39