4 General operating procedures
4.1 Switching the unit on
A
60
4.1.2

Switching the "GALILEOS Compact" on

1. Turn the main switch (A) to position I.
Switching the "GALILEOS Compact" on
2. Wait for approx. 1 minute.
The X-ray radiation indicator (B) lights up for approx. 1 second as
B
a functional check.
After approx. 2 seconds, the green LED (C) in the upper part of
the control panel lights up. This LED remains lit as long as the
unit is on.
C
3. Press the R key.
The unit moves to its starting position.
4. Switch on the PC.
5. Start SIDEXIS XG.
Help message H401 remains displayed on the Multipad as long
as there is no connection with SIDEXIS XG.
4.1.3

Factory setting after switch-on

Factory setting after switch-on
Factory setting after switch-on: GALILEOS
The unit has the following factory configuration on delivery:
● Start settings:
Starting position: from the front (right)
VO1 (for "GALILEOS Comfort")
V04 (for "GALILEOS Compact")
Patient symbol 2: 85 kV/21mAs
● The acoustic signal for end of exposure is activated.
For "GALILEOS Comfort" only:
● The unit language is preconfigured as ordered.
● The welcome screen is switched on.
● The first name, last name and date of birth lines are displayed on the
welcome screen.
If the customer requires a different configuration, this can be implemented
via service routine S017 [ → 252].
Sirona Dental Systems GmbH
Service Manual GALILEOS
61 25 665 D3437
D3437.076.01.15.02 06.2012