Adjusting The Pan Sensor (Xg Plus / 3D Ready / Xg 5 / 3 / 3 Ppe) - Sirona ORTHOPHOS XG 3Dready Service Manual

Table of Contents

Advertisement

4.3 Adjusting the panoramic X-ray unit
Preparations
O
- sensor adjustment
PENING PAN
4 – 12
4.3.3
Adjusting the pan sensor
Plus
ready
(XG
/ 3D
Menu: Diaphragm/system adjustment
Plug the sensor into the sensor slot on the panoramic X-ray unit.
Insert the needle phantom in the bite block holder of the panoramic X-ray
unit (see page 4-9).
ORTHOPHOS XG 3 / 3 PPE:
The fixed diaphragm must be dismantled prior to the sensor adjustment
for the ORTHOPHOS XG 3 / 3 PPE system version (see page 4-10).
-
1.
Open the
PAN
SENSOR ADJUSTMENT
i
NOTE
The menu provides a precision adjustment and a coarse adjustment (preci-
sion adjustment is preset). Perform a precision adjustment first. In most cases,
previous coarse adjustment is not necessary.
/ XG 5 / 3 / 3 PPE)
Pan - Sensor adjustment
submenu (see section 4.2).
D3352.076.01.26.02 09.2016
Tab 4
59 38 399 D3352

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents