4 Installation and start-up
4.3 Moving the sensor
36
4.3

Moving the sensor

If the device is operated using a single sensor, the sensor has to be
moved to the socket on the panoramic rotating unit or the socket on the
cephalometer, depending on the type of exposure being taken.
Note that only one CEPH sensor can be operated at both sockets. The
active sensor area of a PAN sensor is too small for cephalometric images.
Tip: If two sensors (a PAN sensor and a CEPH sensor) are used for
operation, replugging is not required.
NOTICE
When removing the sensor, it can be damaged by impact or if dropped.
The sensor contains an integrated vibration sensor to detect impacts or
falls. If the vibration sensor has triggered, guarantee claims become
void.
➢ Do not drop the sensor under any circumstances!
NOTICE
Electrostatic charges from persons are discharged on the unit.
Electrical components of the unit are destroyed.
➢ Do not touch any electrical components or unprotected plug
contacts.
➢ Discharge yourself by touching a conductive grounded object.
Removing the sensor
1. Hold the sensor firmly.
2. Press the button all the way in and hold it.
The sensor is released from the fastening.
3. Pull the sensor downwards out of the guide.
Sirona Dental Systems GmbH

Operating Instructions ORTHOPHOS XG 5 / Ceph

D3352.201.04.25.23 12.2017
60 04 811 D3352