Sensor Adjustment - Sirona GALILEOS Service Manual

Hide thumbs Also See for GALILEOS:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual GALILEOS
61 25 665 D3437
D3437.076.01.15.02 06.2012
7.3.4

Sensor adjustment

Sensor adjustment
"Service functions"
The
"Sensor
1. Click the
The corresponding tab card is selected.
2. Click on the
SIDEXIS XG makes the unit ready for exposure [ → 162].
Service routine S010.8 is displayed on the control panel.
3. Press the R key to move the unit back to the starting position.
4. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).
IMPORTANT: This process takes approx. 2-3 minutes.
The unit transfers the acquired images to the Reconstruction and
Control Unit (RCU). This process can take 2-3 minutes. Once the
transfer is complete, the evaluation of the sensor calibration is
displayed in the message window.
If the information in the message window indicates that
calibration is
procedure starting with Step b) until calibration is OK and/or
possible.
If you have repeated the procedure three times and still have not
attained a positive result, please contact the SIRONA Customer
Service Center (CSC).
5. If the adjustment is OK or possible, click the
The adjustment is saved.
The sensor adjustment is now complete.
6. Continue the calibration procedure with the iris adjustment [ → 182].
7.3 Adjustment and calibration via the "Service Functions" menu
menu is called [ → 161].
"
tab.
"Image acquisition"
button.
not OK and/or not possible
7 Adjusting/calibrating the unit
, keep repeating the
"Save values"
button.
181

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents