S032: Sensor Test; S032: Test Step 10 - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
SR*
S032
S032.10
S032.50
S032.56
63 03 510 D3352
D3352.076.03.11.02 09.2016
8.18

S032: Sensor test

SA**
Function
Test function for sensor
No
Signal test for sensor in pan slot (combisensor)
No
Self-test of 3D image data path
No
Temperature sensor test, DX88
8.18.1

S032: Test step 10

Signal tests for sensor in pan slot (combisensor)
Selection field
1
1. Call service routine S032.10.
8 Perform service routines via the control panel
Parameters
HSIPan3x3
HSIPan4x4
After the service routine is selected, selection field 1 displays the
current selection for synchronized readout sequence (see table).
8.18 S032: Sensor test
Function
Sensor in Pan mode with
synchronized readout sequence 3x3
clocked out
Sensor in Pan mode with
synchronized readout sequence 4x4
clocked out
341