4 Operation
4.12 X-ray tube unit
170
4.12

X-ray tube unit

The dental treatment unit can be equipped with the HELIODENT
ray tube unit.
The X-ray tube unit is attached to the light post of the dental treatment unit
using a multi-jointed support arm that can be adjusted both horizontally
and vertically. Brakes in the support arm hold the X-ray tube unit in the
position to which it has been adjusted.
The X-ray parameters are set on the wall adapter of the X-ray tube unit or
on a "remote timer".
The intraoral X-ray system XIOS XG USB module can be connected to
the PC via the UBS interface on the dentist element. For details, please
refer to the section "USB interface" [ → 185].
X-ray tube unit, collision while moving the patient chair
WARNING
The X-ray tube unit can be positioned within the movement range of the
patient chair.
Moving the patient chair may cause the patient to collide with the X-ray
tube unit or its support arm. The patient could be injured.
➢ Before moving the patient chair, position the X-ray tube unit to make
a collision with the patient or the patient chair impossible.
Support arm of the X-ray tube unit, risk of crushing in the hinge gaps
CAUTION
Gaps appear between the internal hinges when moving the support
arm.
Fingers may be crushed in these gaps.
➢ Ensure that you never place your fingers in the gaps between the
hinges, neither during operation nor for cleaning purposes.
IMPORTANT
See also the HELIODENT
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
PLUS
operating instructions.
D3509.201.03.03.23 10.2016
PLUS
X-
65 48 510 D3509