Service Manual GALILEOS
SR*
SA**
S005
S005.1
No
Yes
S005.4
No
S005.5
No
S005.8
Yes
61 25 665 D3437
D3437.076.01.15.02 06.2012
8.5

S005: General X-ray tube assembly service

S005: General X-ray tube assembly service
Function
General X-ray tube assembly service
If tube type is valid:
Read out tube type
If tube type is not valid:
Select tube type
Fan test
Temperature sensor test, single tank
Automatic adjustment of pulse preheating
* SR=service routine, ** SHZ=security access
8.5.1

S005: Test step 1

S005: Test step 1
Read or select the X-ray tube assembly type
Tube type GALILEOS
IMPORTANT
Only CB tube D151 R (indicator number 03) is permissible for use with
the GALILEOS volume tomography unit.
➢ Call service routine S005.1 [ → 213].
Selection field 1 shows the indicator number of the installed tube
type.
IMPORTANT
The X-ray tube assembly automatically queries the tube type
information. If no defined value is saved, the security access will be
shown instead of the tube type.
S005.1 GALILEOS
8.5 S005: General X-ray tube assembly service
8 Service
223