Safety-Related Checks By The Service Engineer In Compliance With Standard Iec 62353; General Information For The Service Engineer - Sirona ORTHOPHOS XG 5 Maintenance Manual

Hide thumbs Also See for ORTHOPHOS XG 5:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
3
Safety-related checks by the
service engineer in compliance with
standard IEC 62353
3.1

General information for the service engineer

The Service Manual for ORTHOPHOS XG (order no. 59 38 399) is absolutely essential for performing safety-
related checks, as the protective ground wire and device leakage current tests are described in this document.
Measurements
Switch the power OFF at the main switch of the building installation.
Select the correct current/voltage type and adjust the measuring range to match the expected readings.
Perform continuity tests only on units which are switched off.
ATTENTION
When opening the unit:
Please observe the usual precautionary measures for handling printed circuit boards (ESD).
Touch a ground point to discharge static electricity before handling any components.
ATTENTION
If the dental treatment center does not pass the safety tests, it must not be operated any longer!
You must advise the user of this fact in your capacity as service engineer. Corresponding repair work by
an authorized service engineer is required before putting the system into service again.
NOTE
The safety checks correspond to the standard IEC 62353:2007 (DIN EN 62353/VDE 0751-1:2008).
If you use an automatic tester, you can program it according to these standards.
– Application components Type B
– Permanently attached unit
– Protection Class I
The measurements to be performed are complex and time-consuming.
Sirona therefore explicitly recommends using an automatic tester.
59 87 685 D 3352
D 3352.105.01.11.02

3 Safety-related checks by the service engineer in compliance with standard IEC 62353