8 Perform service routines via the control panel
8.6 S005: General X-ray tube assembly service
SR*
S005
S005.1
S005.2
S005.4
S005.5
S005.6
S005.7
If the tube type is invalid ...
272
8.6

S005: General X-ray tube assembly service

SA**
Function
X-ray tube assembly service XG 3D
General X-ray tube assembly service
If tube type is valid:
No
Read out tube type
If tube type is not valid:
Yes
Select tube type
Yes
Automatic adjustment of preheating
No
Fan test
No
Temperature sensor test, single tank
Yes
Switch off cool-down interval of the single tank
No
Configuring continuous operation of the tube assembly fan
* SR=service routine, ** SHZ=security access
8.6.1

S005: Test step 1

Read or select the X-ray tube assembly type
IMPORTANT
Tube type XG 3D
Only Siemens tube 90/15 (code 04) is valid for operation with the
ORTHOPHOS XG 3D volume tomograph!
Selection field
1
* Factory setting
➢ Call service routine S005.1.
IMPORTANT
The X-ray tube assembly automatically queries the tube type
information. If no defined value is saved, the security access will be
shown instead of the tube type.
1.
S005.1 XG 3D
Service Manual ORTHOPHOS XG 3D / Ceph
Code
Tube type
00
Siemens tube 90/15
01
CEI OPX 100
02
Toshiba tube DO 56
03
CB tube D 151 R
04
Siemens tube 90/15 in 3D single tank*
Selection field 1 shows the indicator number of the installed tube
type.
Confirm the security access.
Selection field 1 shows the indicator number of the (invalid) tube
type detected by the X-ray tube assembly.
Sirona Dental Systems GmbH
63 03 510 D3352
D3352.076.03.11.02 09.2016