4.7 Water unit
158
4.7

Water unit

4.7.1

Swiveling the cuspidor bowl

The cuspidor on the water unit can be manually swiveled approx. 110 mm
toward the patient chair.
A setting can be made that causes the cuspidor bowl to automatically
Reference "Link cuspidor to mouth rinsing position"
move inward when the mouth rinsing position chair program (S) is
selected; see"Linking the movement of the cuspidor bowl to the mouth
rinsing position" [ → 191].
IMPORTANT
Cuspidor, motor-driven, starting position
The cuspidor bowl automatically swivels back to its starting position
The cuspidor bowl automatically returns beforehand to ensure that the
patient does not collide with it during chair movements. This return travel
is dependent on the chair movement and is executed only if a collision
hazard exists.
4.7.2

Tumbler filling with automatic sensor control

The automatic sensor control is not available in all countries.
Not available in all countries.
With this tumbler filler, the automatic sensor control detects the filling
level of the tumbler automatically.
If the tumbler is removed before the preset filling level has been reached,
the water flow stops immediately.
Filling the tumbler
The tumbler must not be transparent.
The tumbler must always be placed in the center of the area provided
for it.
➢ Place the tumbler under the tumbler filler.
 The tumbler is filled automatically.
 After the preset filling level has been reached, the water flow stops
automatically.
If necessary, the tumbler can be filled manually with the
Cross-reference to TENEO
key, see "Tumbler filling" [ → 80].
IMPORTANT
Automatic sensor control after switch-on
Tumbler filling after switch-on
If an empty tumbler is standing below the tumbler filler when the dental
treatment unit is switched on, the tumbler will not be filled automatically.
To activate automatic tumbler filling, remove the tumbler briefly and then
put it back again.