7.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
7.2
Adjustment and calibration via the "Unit
adjustment/calibration" menu
IMPORTANT
After every adjustment or calibration of the unit, the reference values for
the constancy measurement must be recalculated and entered in the
"Test results" form, "Reference value" column.
7.2.1

2D adjustment

Pan sensor adjustment

7.2.1.1
7.2.1.1.1
Automatic adjustment: Panoramic sensor (standard)
The pan sensor is plugged into the pan slot on the unit.
The OP XG needle phantom is in the pan bite block holder on the unit
[ → 198].
1. Call the
"Adjustment/calibration"
2. In the structure tree, under
Sensor Adjustment"
The
area. The menu supports precision adjustment and coarse
adjustment (precision adjustment is pre-selected). Perform a
precision adjustment first. In most cases, previous coarse
adjustment is not necessary.
3. Make SIDEXIS XG ready for exposure. [ → 197]
4. Take an exposure (60 kV / 3 mA; 0.6 s) [ → 197].
SIDEXIS XG calculates the adjustment values S1, S2, and S3
automatically from the exposure and enters them in the text
boxes of the
The exposure is displayed in the exposure window in SIDEXIS
XG.
menu [ → 188].
"2D Adjustment"
element.
"Pan - Sensor Adjustment"
"Pan - Sensor Adjustment"
7 Adjusting and calibrating the unit
"Pan -
, click on the
menu is displayed in the action
menu.