7.1
General information about unit adjustment and
calibration
Checking the mechanical unit adjustment: GALILEOS

General information about unit adjustment and calibration

Start by checking the mechanical unit adjustment. This step is a
prerequisite for the subsequent adjustment and calibration of the unit.
Please adhere to the following order when adjusting and calibrating the
system:
GALILEOS calibration procedure
● Diaphragm exposure: "Type 1/Type 2" [ → 167] diaphragm or
diaphragm exposure: "Type 3" diaphragm [ → 171]
● Checking the radiation field [ → 178]
● Dosimetry [ → 179]
● Sensor adjustment [ → 181]
● Iris adjustment [ → 182]
● Shading calibration [ → 183]
● Distortion calibration [ → 185]
● Geometry calibration [ → 187])
Tip: It may be helpful to use the SIDEXIS XG coloring function to evaluate
the image.