4 Electrical connection

4.1 Connecting the control cables (PAN)

4.1 Connecting the control cables (PAN)
L117
L25
Duplex patch cable,
1:1 connection,
SC/SC 50/125 μm
1. Connect the personal computer to socket SC:SC of the
panoramic X-ray unit via the media converter (see the
Operating Instructions supplied with the media convert -
er).
Install the media converter at a suitable location using
fastening screws or the Velcro strap supplied for this
purpose.
NOTICE
Do not fasten the media converter to the X-ray unit.
*) not included in the scope of supply
46
X108
X101
3.
2.
Switch*
Cat.5
RJ45
Installation Instructions ORTHOPHOS XG 3
L117
Coiled cable
Media convert -
L25
For the installation version without remote control
2.
Connect the coiled cable of the release button to socket
X103 of the panoramic X-ray unit.
For the installation version with remote control
Connect the remote control to socket X103 of the pan -
3.
oramic X-ray unit with cable L117.
Secure the connector on socket X103.
Sirona Dental Systems GmbH
1.
60 51 416 D 3352