S017: Test step 15
11.10.3.11
Selection field
1
* Factory setting
Activate/deactivate the acoustic signal for end of exposure
1. Call service routine S017.15 [ → 125].
S017_5 XG 3D
S017_5 GALILEOS / XG 3D / XG 3D ready
 Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
2. Use the arrow keys (A) to select the required setting in selection field
1 (see table).
 Once the required setting has been selected, the Memory key
lights up.
3. Save the setting [ → 129].
4. Exit the service routine [ → 129].
11.10 Service routines for startup
Code
Function
00
Acoustic signal indicating the end of the
exposure is disabled
01
Acoustic signal indicating the end of the
exposure is enabled*
11 Initial startup