Software Update; Service Menu - Sirona SIROLaser Xtend Operating Instructions Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Operating Instructions SIROLaser Xtend
63 18 351 D3576
D3576.201.01.05.02    11.2018
NOTE
The power limit directly influences the number of applications that can
be used by this user.
For example, when you choose a power limit of 2 W the user can not
choose preset indication with more than 2 W. If the power limit 0.5 W
(default) so the user has no access to preset indications.
The key user is able to select the applications this user is allowed to
use.
NOTE
The non-usable applications due to the direct selection or the power
limit appear shaded and are disabled.
5.5.4.6

Software update

If a software update of the SIROLaser Xtend is needed please proceed
as follows:
1. Choose in main menu the item "Set-up".
2. Choose there the item "Software update" (these can be found on
the second sheet).
3. Follow the instructions in the note message and insert the USB
stick.
NOTE
Please bear in mind that the system requires approx. 5 seconds to
detect the USB stick.
4. Press "OK".
Ä It appears the note message with the information that the
software update is in progress.
Ä The note message disappears, the start screen appears and
the self test is in progress.
CAUTION
Leave the USB stick and the power cable plugged in until the
completion of software update.
The Software update may take up to five minutes.
5. Please now, like when you turn on the device, press 'OK'.
Ä The software update was successfully performed. The USB
stick can be removed.
5.5.4.7

Service Menu

NOTE
Only authorized persons are allowed to enter the service menu. To
avoid misuse it is necessary to enter the eight-digit pin code.