5 Messages
5.5 List of error messages
Error code
Description
E7 15 01
Wrong memory modules.
Error code
Description
E7 15 03
Wrong software constellation of
modules.
Error code
Description
E6 15 04
Product activation keys invalid or
not available.
Occurs after replacement of tube
assembly (DX6) or board DX11 and
possibly after software updates.
See also the section titled Measures
following replacement of
boards [ → 347].
Error code
Description
E6 15 05
Unit serial number invalid or not
available.
Occurs during first power-on after
replacement of board DX6 or DX11.
See also the section titled Measures
following replacement of
boards [ → 347].
Error code
Description
E6 15 10
Update file for module is corrupt.
106
5.5.7

Location 15: Configuration, update

Location 15: Configuration, update
Actions required
If a DRAM memory module is plugged into
board DX11 ...
● Replace memory module or DX11.
If no DRAM memory module is plugged into
board DX11 ...
● Replace board DX11.
E7 15 03
Actions required
● Check the software versions of the unit (on
the info screen or by running service
routing S008.2), and run or repeat
software update or downgrade if
necessary.
E6 15 04
Actions required
● Enter release key.
*) OI = Operating instructions
E6 15 05
Actions required
● Run service routine S008.3 and confirm or
enter the unit serial number on the unit.
E6 15 10
Actions required
● Obtain latest update file from the Sirona
Customer Service Center (CSC) or the
Sirona home page and perform software
update.
Sirona Dental Systems GmbH
Service Manual GALILEOS
see
S. [ → 350]
S. [ → 350]
see
S. [ → 238],
S. [ → 61]
see
see OI*
see
S. [ → 238]
see
S. [ → 61]
61 25 665 D3437
D3437.076.01.15.02 06.2012