Measurement Examples
Measuring Parameters with Cable
5. Setting the Time Domain Function
Set the conversion function to display the response in time the domain. If you enable this
setting, the response in time domain is displayed as shown in Figure 15-34. A peak
indicating a small mismatch appears at the location of the connector.
Setting Description
Data format: real
Setting the transformation type to low-pass
impulse
Setting the window type to maximum.
Setting the display range: from 0 s to 10 ns
Enabling the transformation function
Executing auto scale
Figure 15-34
Response in time domain
520
Key Operation
- Real
- Transform - Lowpass Imp.
Window - Maximum
Start -
Stop -
Transform (set to ON)
- Auto Scale
Chapter 15