5. Through Calibration (THRU CAL)

Required Equipment

Attenuator Pad 20dB
N(f)-BNC(m) Adapter
N(m)-BNC(f) Adapter
BNC(m)-BNC(m) Cable, 30 cm
Power Splitter

Procedure

1. Connect the equipment as shown in Figure 2-7.
2. Select \THRU CAL" and click
3. Following the program instructions, complete the procedure.
2-14 Adjustments and Performance Tests
8491A Opt.020
PN 1250-1477
PN 1250-1476,
E4915A: 1ea, E4916A: 4ea
PN 8120-1838,
E4915A: 1ea, E4916A: 3ea
11667A (E4916A)
Figure 2-7. Through Calibration Setup
NNNNNNNN
OK in the list box of the program display.