CAL ON U
appropriate measurement.

View Cal

From the Choose Calibration screen (page 66), press
following screen:
The top box shows the properties of the current calibration that you performed
and the displayed S-parameters that it is correcting.
The bottom box shows the S-parameters that are displayed but NOT corrected by
the current performed calibration, but rather the CalReady calibration. Learn
more about CalReady on page 65.

Calibration Type

The FieldFox simplifies the calibration process by recommending the most
accurate and efficient calibration type based on the displayed S-parameters.
However, there may be times when you may want a little more accuracy or a
little faster sweep time. The following information can help you learn about the
various calibration choices.
Definitions:
Non-insertable DUT – A device whose
connectors could NOT mate together.
They either do not have the same type
of connector or they have the same
gender. This also means that the test
port cables could NOT mate together as
in the above diagram.
Calibration for NA, CAT, and VVM Modes
is shown on the screen when a Response Cal is correcting ONLY the
to see the
View Cal
Insertable DUT – A device whose
connectors could mate together. They
have the same type of connector and
opposite or no gender. This also
means that the test port cables could
mate together, as in the above
diagram.
73