Calibration

THRU Response Calibration (Transmission Test)

THRU Response Calibration (Transmission Test)
In THRU response calibration, calibration data are measured by connecting a THRU
standard to the desired test port. This calibration effectively eliminates the frequency
response transmission tracking error from the test setup in a transmission test using that
port (Figure 4-6). It is also possible to carry out an isolation calibration using a LOAD
standard in the process of THRU response calibration. An isolation calibration will
eliminate isolation error (crosstalk error) from the test setup in a transmission test using
that port.
Figure 4-6
2-Port Error Model (Thru Response)
Figure 4-7
2-Port Error Model (Thru Response + Isolation)

Procedure

Step 1. Press
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Select the test ports (and corresponding S parameters) upon which a THRU response
calibration will be performed.
Softkey
2-1 (S21)
3-1 (S31)
88
.
.
Calibrate
.
Response (Thru)
.
Select Ports
Function
Selects test port 2 (input) and test port 1 (output). Corresponds to the
determination of S
Selects test port 3 (input) and test port 1 (output). Corresponds to the
determination of S
.
21
.
31
Chapter 4