One-Port Reflection Error Correction - Agilent Technologies 8753ET User Manual

Network
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Calibrating for Increased Measurement Accuracy

One-Port Reflection Error Correction

One-Port Reflection Error Correction
• removes directivity errors of the test setup
• removes source match errors of the test setup
• removes frequency response of the test setup
You can perform a 1-port correction for an S
between the two procedures is the measurement parameter that you select.
NOTE
This is the recommended error correction process for all reflection measurements, when full
two-port correction or enhanced response calibration is not used.
1. Press
.
Preset
2. Select the type of measurement you want to make.
❏ If you want to make a reflection measurement on PORT 1 (in the forward direction, S
instrument default setting.
❏ For ES analyzers, if you want to make a reflection measurement on PORT 2 (in the reverse direction,
S
), press:
22
Meas
Refl: REV S22 (B/R)
3. Set any other measurement parameters that you want for the device measurement: power, number of
points, IF bandwidth.
4. To access the measurement correction menus, press:
Cal
5. If your calibration kit is different than the kit specified under the
CAL KIT
SELECT CAL KIT
If your type of calibration kit is not listed in the displayed menu, refer to
page
7-57.
6. To select the correction type, press
❏ If you want to make a reflection measurement at PORT 1, press:
S11 1-PORT
❏ For ES analyzers, if you want to make a reflection measurement at PORT 2, press:
S22 1-PORT
7. Connect a shielded open circuit to PORT 1 (or PORT 2 for an S
NOTE
Include any adapters that you will have in the device measurement. That is, connect the
calibration standard to the particular connector where you will connect your device under
6-26
or an S
measurement for ES analyzers. The only difference
11
22
(select your type of kit)
CALIBRATE MENU
softkey, press:
CAL KIT [ ]
RETURN
"Modifying Calibration Kits" on
and select the correction type.
measurement).
22
), leave the
11