Measurement Support Functions; Error Correction Function; Dc Resistance (Rdc) Measurement - Agilent Technologies Twelfth 4287A RF LCR Operation Manual

Table of Contents

Advertisement

Specifications and Supplemental Performance Characteristics

Measurement Support Functions

Measurement Support Functions

Error correction function

Available Calibration and Compensation
OPEN/SHORT/LOAD
Calibration
Low-Loss Capacitor Calibration
Port Extension Compensation
(Fixture Selection)
OPEN/SHORT Compensation
Calibration/Compensation Data Measurement Point
Data Measurement Points

DC resistance (Rdc) measurement

Measurement range
Measurement resolution
Test signal level
Error correction
Measurement uncertainty
270
Connect OPEN, SHORT, and LOAD standards to the
desired reference plane and measure each kind of
calibration data. The reference plane is called the calibration
reference plane.
Connect the dedicated standard (Low-Loss Capacitor) to the
calibration reference plane and measure the calibration data.
When a device is connected to a terminal that is extended
from the calibration reference plane, set the electrical length
between the calibration plane and the device contact. Select
the model number of the registered test fixtures in the
4287A's softkey menu or enter the electrical length for the
user's test fixture.
When a device is connected to a terminal that is extended
from the calibration reference plane, make OPEN and
SHORT states at the device contact and measure each kind
of compensation data.
Same as measurement points set in the measurement point
setup display. (Changing the frequency or oscillator level
settings after calibration or compensation makes the
calibration and compensation data invalid.)
0.1 Ω to 100 Ω
1 mΩ
1 mA (maximum)
OPEN/SHORT/LOAD Calibration, OPEN/SHORT
Compensation. (Changing the frequency or oscillator level
settings after the calibration or compensation makes the
calibration and compensation data invalid.)
R
0.05
dut
±
×
--------- -
-------------- -
1
+
+
100
R
10000
dut
Rdut: DC resistance measurement value [Ω]
Within ±5°C from the calibration temperature.
Measurement accuracy applies when the calibration is
performed at 23°C ±5°C. When the calibration is performed
beyond 23°C ±5°C, the measurement accuracy decreases to
half that described.
[%]