Full 2-Port Calibration Using The 2-Port Ecal Module - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Calibration
ECal (electronic calibration)
Step 3. Press
calibration.
Step 4. Press
Step 5. Press
Step 6. Press
Step 7. Perform a 1-port calibration.
Softkey
Port 1
Port 2
Port 3
Port 4

Full 2-Port Calibration Using the 2-Port ECal Module

Follow the procedure below to perform a full 2-port calibration using the 2-port ECal
module.
Step 1. Connect the USB port on the ECal module with the USB port on the E5070B/E5071B via a
USB cable. This connection may be done while the E5070B/E5071B's power is on.
Step 2. Connect port A and port B on the ECal module to the test ports to be calibrated.
NOTE
You can connect the ports of the ECal and the test ports of the E5070B/E5071B
arbitrarily.Connected ports can be manually specified although they are automatically
detected before data measurement. For more information, "Turning off ECal auto-detect
function" on page 131.
128
or
.
.
ECal
.
1 Port ECal
Function
Performs a 1-port calibration on test port 1
Performs a 1-port calibration on test port 2
Performs a 1-port calibration on test port 3
*1
Performs a 1-port calibration on test port 4
*2
*1.Only with Options 313, 314, 413, and 414.
*2.Only with Options 413 and 414.
to select the channel for which you want to perform the
Chapter 4