Calibrating for Increased Measurement Accuracy
Frequency Response Error Corrections
Frequency Response Error Corrections
You can remove the frequency response of the test setup for the following measurements:
• reflection measurements
• transmission measurements
• combined reflection and transmission measurements
Response Error Correction for Reflection Measurements
1. Press
.
Preset
2. Select the type of measurement you want to make.
If you want to make a reflection measurement on PORT 1 (in the forward direction,
S
), leave the instrument default setting.
11
For ES analyzers, if you want to make a reflection measurement on PORT 2 (in the
reverse direction, S
Refl: REV S22 (B/R)
Meas
3. Set any other measurement parameters that you want for the device measurement:
power, sweep type, number of points, or IF bandwidth.
4. To access the measurement error-correction menus, press:
Cal
CALIBRATE MENU
5. If your calibration kit is different than the kit specified under the
softkey, press:
CAL KIT
Cal
If your type of calibration kit is not listed in the displayed menu, refer to
Calibration Kits" on page
6. To select a response correction, press:
Cal
CALIBRATE MENU
Connect the short or open calibration standard to the port you selected for the test port
(PORT 1 for S
or PORT 2 for S
11
NOTE
Include any adapters or cables that you will have in the device measurement.
That is, connect the standard device to the particular connector where you
will connect your DUT.
6-12
), press:
22
(select your type of kit)
SELECT CAL KIT
7-57.
RESPONSE
on ES analyzers).
22
CAL KIT [ ]
Cal
RETURN
"Modifying