Sirona Dental Systems GmbH
Service Manual GALILEOS
61 25 665 D3437
D3437.076.01.15.02 06.2012
4.2
Updating the firmware
Updating the firmware
4.2.1
Updating the unit firmware
GALILEOS software update
Updating the unit firmware
Read the information provided on the software CD supplied with the unit
and on the SIRONA dealer page on the Internet very carefully. These
sources always contain the latest information on software updates.
"SIDEXIS Manager"
1. Start the
"SIDEXIS XG"
"Configuration of the X-ray components"
2. Click on
"Configuration of the X-ray components"
The
3. Select the
"Attributes"
4. Click on the
The dialog box for entering the service password opens.
5. Enter the service password.
Enter the first 4 digits of the current system date in reverse order as
the service password (e.g. on 05/24/1995, 5042 must be entered as
the service password.
If an incorrect service password or no password at all is entered, the
limited update menu for users will be started. This only supports an
automatic update option.
The dialog box for selecting the installation source opens.
6. Click on the button with the 3 dots.
"imagefile" is the default installation source for the software update.
under "Start"/"Programs"/"SIDEXIS"/
.
tab.
"Software update"
button.
4 General operating procedures
4.2 Updating the firmware
.
menu opens.
61