4 General operating procedures
4.2 Updating the unit firmware
62
4.2

Updating the unit firmware

IMPORTANT
Downgrading to older versions
Downgrading the unit software version V04.14.00 or higher to an older
version is not a simple process. However, if this is essential, please get
in contact with the SIRONA Customer Service Center (CSC) in
advance.
Read the information provided on the firmware CD supplied with the unit
XG 3D software update
and on the SIRONA dealer page on the Internet very carefully. These
sources always contain the latest information on software updates.
"SIDEXIS Manager"
1. Start the
"SIDEXIS XG"
"Configuration of the X-ray components"
2. Click
"Configuration of the X-ray components"
The
"Attributes"
3. Select the
4. Click on the
The dialog box for entering the service password opens.
5. Enter the service password.
Enter the first 4 digits of the current system date in reverse order as
the service password (e.g. on 05/24/1995, 5042 must be entered as
the service password.
If an incorrect service password or no password at all is entered, the
limited update menu for users will be started. This includes only the
possibility for an automatic update [ → 67].
The dialog box for selecting the installation source opens.
Service Manual ORTHOPHOS XG 3D / Ceph
under "Start"/"Programs"/"SIDEXIS"/
.
tab.
"Software update"
button.
Sirona Dental Systems GmbH
.
menu opens.
63 03 510 D3352
D3352.076.03.11.02 09.2016