Diaphragm Test Exposures - Sirona ORTHOPHOS XG 3Dready Service Manual

Table of Contents

Advertisement

Tab 7
59 38 399 D3352
D3352.076.01.26.02 09.2016
7.9.2

Diaphragm test exposures

The diaphragm test exposures are used for a quick check of the system
adjustment. The
TEST EXPOSURE
word. Diaphragm test exposures can thus be performed by the user as well.
Menu: Test exposure
The
menu is started from SIDEXIS XG:
TEST EXPOSURE
U
C
TILITIES
ONSTANCY TEST
D
VICE EXPOSURE
IAPHRAGM TEST EXPOSURE
i
NOTE
When you open the
TEST EXPOSURE
mode to the PC service mode logged by the PC.
In the PC service mode, the control options that are available on the Easypad
are determined by SIDEXIS and the currently selected service routine. Gen-
eral control of the unit (as in the user mode) is not possible in this mode.
By clicking the corresponding tab with the mouse, you can switch between the
,
PAN DIAPHRAGM
PAN SYMMETRY
menus. To quit the
TSA SENSOR
i
NOTE
The Ceph - Fixed point of rotation and TSA - Sensor tabs are not present in the
XG 3 / 3 PPE system version.
7.9 Checking the diaphragm
menu is started without a service pass-
S
X-
XCXP
ELECT
RAY DEVICE
menu, the unit switches from the user
,
CEPH FIXED POINT OF ROTATION
menu, click
TEST EXPOSURE
S
-
ER
, and
.
CANCEL
7 – 21