8 Perform service routines via the control panel
8.19 S033: Test of ceph image path without SIDEXIS XG (for ceph units only)
SR*
S033
S033.10
346
8.19
S033: Test of ceph image path without SIDEXIS
XG (for ceph units only)
SHZ**
Function
Test of ceph image path without SIDEXIS XG
No
Signal tests for sensor in ceph slot
* SR=service routine, ** SHZ=security access
8.19.1

S033: Test step 10

Signal tests for sensor in ceph slot
Selection field
1
1. Call service routine S033.10 [ → 262].
2. Use the arrow keys (A) in selection field 1 to select the mode of the
required synchronized readout sequence (see table).
IMPORTANT
If the R key should light up first, the sensor holder must first be moved
to the corresponding position. To do so, press the R key.
Service Manual ORTHOPHOS XG 3D / Ceph
Parameters
HSIC4x4
HSIC4x4M
After test step 10 is selected, selection field 1 displays the current
selection for synchronized readout sequence.
Once the mode for the synchronized readout sequence has been
selected, the T key (B) lights up.
Sirona Dental Systems GmbH
Function
Sensor in ceph mode with
synchronized readout sequence 4x4
clocked out without ceph scan
Sensor in ceph mode with
synchronized readout sequence 4x4
clocked out with ceph scan
63 03 510 D3352
D3352.076.03.11.02 09.2016