9 Repair
9.11 X-ray tube unit
410
9.11.3
What has to be done after replacing the X-ray tube
assembly?
IMPORTANT
The tube assembly may be replaced only as a complete module.
IMPORTANT
Note the dependency of the tube assembly on device serial numbers in
the spare parts list.
IMPORTANT
Use the most current software version available on the ORTHOPHOS
XG3D CD.
1. Switch the unit on.
NOTICE! Do not acknowledge any error messages at this time.
2. If the software status of the newly installed DX6 board is not
compatible with the current overall system software version, perform
a software update to the current system version.
The current update file is located on the ORTHOPHOS XG3D
Firmware CD. It is delivered with each DX6 replacement board and
also included in the country set. The contents of the CD can be
downloaded from the Dealer domain of the Sirona Internet home
page (under Product Info/X-ray Systems): www.sirona.com
3. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
Error message E6 11 07 (undefined system class) is displayed.
4. Acknowledge the error message with the R key.
The access level for the service menu (level 4) is automatically
started.
5. Press and hold down the service key (D) until the patient symbol keys
light up (approx. 2 s).
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352
D3352.076.03.11.02 09.2016