9.6.3
What has to be done after replacing the X-ray tube
assembly?

What has to be done after replacing the X-ray tube assembly?

IMPORTANT
Since board DX6 is also replaced with the tube assembly, you MUST
also follow the instructions in the chapter entitled Measures following
replacement of boards [ → 347].
1. Perform a complete unit adjustment or calibration [ → 155].
2. Perform an acceptance test (for Germany only) without calling in an
expert.