9 Repair
9.10 Diaphragm unit
A
404
9.10

Diaphragm unit

9.10.1

Replacing the diaphragm unit

IMPORTANT
Only the diaphragm unit of ORTHOPHOS XG 3D may be installed! Note
the dependency of the diaphragm unit on device serial numbers in the
spare parts list.
IMPORTANT
The diaphragm unit may be replaced only as a complete module.
IMPORTANT
Importance of the tube assembly version
Depending on the device serial number or tube assembly version, the
tube assembly installed in your device may differ slightly from the figures
in this section.
1. Remove the "Rear tube assembly" cover [ → 41].
2. Pull cables L11 and L12 off connectors X501 (L11) and X101 (L12)
on board DX61.
3. Pull off the cables L21 (2x) and L20 from the connectors X303, X304
(L21) and X203 (L20) on board DX61.
4. Loosen the screw (A) at the top flap of the diaphragm unit and unhook
the defective diaphragm unit from the brass rings (B) at the tube
assembly.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352
D3352.076.03.11.02 09.2016