2.6 List of available service routines

Routine
Description
S007
Error logging
Test step 1:
Display error logging memory
Test step 2:
Delete error logging memory
Test step 5 (V02.27 or higher):
Enabling CAN bus logging in the Mini-
web
S008
Update service
Test step 2:
Display of module software version
statuses
Test step 3:
Input/confirmation of
unit Serial number
Test step 4
(not for XG 3 or XG 3PPE)
Initializing the function activation
S009
Flash file system
Test step 4:
Formatting flash file system
S012
CAN bus service
(from
Test step 2:
V02.27)
Query of CAN status register of mod-
ules
Test step 3:
Reset of CAN status register of mod-
ules
Test step 4:
Display of CAN bus cycle on LEDs of
modules
S014
Rotation motor service
(from
Test step 1:
V02.28.02)
Travel of rotational drive to the Pan
home position
Test step 2:
Travel of rotational drive to the Ceph
position
Test step 3:
Free travel of rotational drive
Test step 4:
Display of light barrier signals of rota-
tional drive
2 – 60
Required...
after error messages
only after consulting with the Sirona Customer
Service Center!
for checking the configuration
after error messages
after error messages
on suspicion of unstable CAN bus
in case of accumulated error messages EX yy
12; Ex yy 06, Ex yy 07, Ex yy 03
not suitable as an accompanying measure e.g.
"check connector of CAN bus" and "check bus
terminations"
to be executed prior to service routine S012.2 if
required
in case of sporadic failures of modules on CAN
bus (after Ex yy 06, Ex yy 07,
Ex yy 08, Ex yy 12)
as accompanying measure when checking con-
nectors (wiggle) for firm seating
for electromechanical problems with the ring
motor or the ring rotation system
Tab 2