Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352
D3352.076.03.11.02 09.2016
8.5 S002: Radiation without rotary movement, selectable kV/mA level and maximum
8.5.2

S002: Test step 3

Radiation without rotary movement; selectable kV/mA level and
maximum radiation time (max. 14 s); primary diaphragm fully opened
with respect to PAN position
WARNING
Unit is radiating X-rays.
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away
from the unit as the coiled cable for the release button allows you to.
Selection field
1
2
Factory settings: 60kV/8mA, 2s
1. Call service routine S002.3.
Calling this service routine moves the primary diaphragm to the
new position (fully opened with respect to PAN position). During
this time, a progress indicator is displayed in selection field 1.
2. Use the arrow keys (A) to select the required kV/mA level and the
required radiation time (see table).
8 Perform service routines via the control panel
Parameters
Range of values
kV/mA level
60kV/3mA –
90kV/12mA –
Radiation time
0.1s – 4s
267