S002: Radiation Without Rotary Movement, Selectable Kv/Ma Level And; S002: Test Step 5 - Sirona GALILEOS Service Manual

Hide thumbs Also See for GALILEOS:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual GALILEOS
SR*
SHZ**
S002
S002.5
Yes
61 25 665 D3437
D3437.076.01.15.02 06.2012
8.4 S002: Radiation without rotary movement, selectable kV/mA level and maximum radiation time
8.4
S002: Radiation without rotary movement,
selectable kV/mA level and maximum radiation
time
S002: Radiation without rotary movement, selectable kV/mA level and maximum radiation time
Function
X-ray beam test
Long-term exposure with fixed radiation intervals from any position
* SR=service routine, ** SHZ=security access
8.4.1

S002: Test step 5

S002: Test step 5
Long-term exposure with fixed radiation intervals from any position
WARNING
Unit is radiating X-rays
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away
from the unit as the coiled cable for the release button allows you to.
Selection field
1
2
* Factory setting
Parameters
Range of values
kV/mA level
60 kV/8 mA –
85 kV / 7 mA*
Radiation time
0.1 s – 5.0 s*