S002.1: Radiation w/o rot. movement, max. radiation time selectable, prim. diaphr. in last
5.5 Service routine S002
Selecting test step 1
Confirming the security access
Selecting the kV/mA level and radiation
time
5 – 26
5.5.1

S002: Test step 1

1.
Select test step 1 in selection field 2 with the arrow keys and confirm your
selection by pressing the Service key
Plus
ready
XG
/ 3D
:
Easypad touchscreen
1
-
S002
1
-
2
3
2.
Confirm security access by once again selecting the number of the main
routine (2) with the arrow keys of selection field 2 and then pressing the
Service key
.
Plus
ready
XG
/ 3D
:
Easypad touchscreen
1
2
-
2
3
3.
Use the arrow keys of
– Selection field 1 to select the desired kV/mA level and
– Selection field 2 to select the desired radiation time.
Plus
ready
XG
/ 3D
:
Easypad touchscreen
E
X RAY
1
-
60kV/8mA
2.0 s
2
-
3
.
XG 5 / 3 / 3 PPE: Multipad
+
S002_______1____
+
1
XG 5 / 3 / 3 PPE: Multipad
___________2____
+
1
XG 5 / 3 / 3 PPE: Multipad
+
60/8____2.00____
+
1
D3352.076.01.26.02 09.2016
Tab 5
2
3
2
3
2
3
59 38 399 D3352